Infusion Pump ManoPump iV1
Infusion
Infusion Pump ManoPump iV1
Safe and precise administration
Practical, ergonomic and efficient:
- Small and light (1.45 kg)
- Kennel mount and IV stand support included
- High-performance battery (up to 10 hours run time)
- Large touchscreen (4.3’’-10.75 cm) for a clear diplay of parameters including infusion time left
- Many settings available such as:
- Overall infusion volume
- Requested flow
- Infusion total time
- High sensitivity of settings (bubbles and occlusion pressure)
- Transfusion compatible
Application fields for the ManoPump iV1
Intensive care
A continuous infusion is vital for any dehydrated patient.
The flow rate is conditioned by the required volume and the duration of the infusion:
1 - Initial deficit to compensate
Deficit to compensate (liters) = Weight (Kg) x % of dehydration / 100
Example: 3 kg cat dehydrated by 15% -> Q = 3 x 0.15 = 0.45 l (450 ml)
2 - Maintenance need:
Large dog = 60 ml/kg/24h - Medium dog = 50 ml/kg/24h - Small dog or cat = 40 ml/kg/24h
3 - Loss:
4 ml/kg vomiting - 200 ml/kg diarrhea - 1– 2 ml/kg urinating
Surgery
A correctly perfused animal will wake up faster (no hypovolemia), moreover, the kidney failure risk is considerably reduced.
Recommendation for intraoperative infusion flow rates:
Low blood loss: 10ml/kg/h - normal blood loss: 15ml/kg/h - high blood loss: 20ml/kg/h
During postoperative process, the infusion must be maintained until the patient wakes up.
The ManoPump iV1 infusion pump is equipped with a battery running up to 10 hours :
- to monitor the patient before, during and after the surgery
- to allow uninterrupted operation even in case of power shortages
- to be used at the clinic as well as outside (home-based care, transportation)
ManoPump iV1 technical data:
- Accumulated volume from 0.1 to 9999 ml
- Topuchscreen: 4.3'' (10.75 cm)
- Infusion time left
- Flow from 0.1 to 2000 ml/h
- Modes :
- Flow and volume
- Overall time and volume
- Overall time and flow
- Body mass
- Saved modes
- Micro mode
- Accuracy: +/- 5%
- Air bubble detection: 20, 50, 100, 200, 400, 800 µl
- Occlusion settings: 5 levels (50, 150, 300, 600, 900 mmHg)
- Entry capacity: up to 2000
- Data review: 5000
- Quick setting use: access to the 20 latest settings
- Protection classification: IP44
- Dimensions and weight: 13x9x14 cm - 1.45 kg
- Operating life: 10h at 25 ml/h
- Power: 100-240V, 50-60 Hz
- Safety and operation alarms
- Self-diagnostic
- IV stand support and kennel mount support